Accuracy of Signal-to-noise Ratio Measurement Method for Magnetic Resonance Images

Akio Ogura Akira Miyai Fumie Maeda Hiroyuki Fukutake Rikiya Kikumoto

刊行年
2003-01-01
収録
『Japanese Journal of Radiological Technology』 59(4) pp. 508-513
出版
Japan Radiological Society
言語
英語
openalex
W2469144403
doi
10.6009/jjrt.kj00003174111
pmid
12743522
mag
2469144403
issn
0369-4305
URL
https://www.jstage.jst.go.jp/article/jjrt/59/4/59_KJ00003174111/_pdf

要旨

The signal-to-noise ratio (SNR) of a magnetic resonance image is a common measure of imager performance. However, evaluations for the calculation of the SNR use various methods. A problem with measuring SNR is caused by the distortion of noise statistics in commonly used magnitude images. In this study, measurement accuracy was compared among four methods of evaluating SNR according to the size and position of regions of interest (ROIs). The results indicated that the method that used the difference between two images showed the best agreement with the theoretical value. In the method that used a single image, the SNR calculated by using a small size of ROI showed better agreement with the theoretical value because of noise bias and image artifacts. However, in the method that used the difference between two images, a large size of ROI was better in reducing statistical errors. In the same way, the methods that used air noise and air signal were better when applied to a large ROI. In addition, the image subtraction process used to calculate pixel-by-pixel differences in images may reach zero on a minus pixel value when using an image processor with the MRI system and apparatuses associated with it. A revised equation is presented for this case. It is important to understand the characteristics of each method and to choose a suitable method carefully according to the purpose of the study.

主題

この書誌の出所

  • openalex— W2469144403(2026-08-13取得)

引用キー: Ogura2003AccuracySignalNoise

書誌 56,535件 語別索引 17,251件 資源 113件 研究者 303名 JSON